HEMO Control Analyzer

Accurate Hemoglobin and Hematocrit results with one simple test

Sample volume of only 10μl and results within 25 to 60 seconds. Compact in size with no maintenance required and a rechargeable battery which lasts up to 100 hours.

Download Brochure  |  Download User Manual

2 RESULTS IN 1 TEST

STEP 1
Take a blood sample

STEP 2
Insert cuvette into analyzer

STEP 3
Results in ca. 25-60 sec

HEMATOCRIT

  • Hematocrit (Hct) is the volume of red blood cells within a sample.
  • Hematocrit is an essential measurement in emergency and transfusion medicine as well as in therapeutic blood dilution.
  • Hematocrit readings also support the diagnosis of conditions such as anemia and polycythemia.
  • On Hemo Control the Hct value is derived from the hemoglobin result using the factor 2,94.
  • Readings are only provided for normal hemoglobin ranges i.e. 12 – 18 g/dL. Abnormal values will not be reported.

PRACTICAL & PORTABLE

  • Hemoglobin and hematocrit results from one sample available after 25–60 seconds (depending on the concentration)
  • Venous, arterial or capillary blood
  • Sample volume only 8 or 10 µl depending on cuvette type
  • n•x•t shape microcuvette minimises air bubbles
  • Compact in size and weighing just 700g
  • No maintenance required. Auto self-test
  • Integrated rechargeable battery (100 hours)

EKF NXT MICROCUVETTE

Unique design minimises risk of air bubbles in the sample:

  • A slot at the rear of the sample chamber allows trapped air to escape
  • Round tip and wide opening for sample uptake
  • Users can hold the cuvette at any angle to the blood droplet
  • Desiccant conveniently integrated in canister
  • Total shelf-life up to 24 months, 3 months after opening of canister
  • Package configuration 50 pcs per canister (4x50pcs)

Note: Cuvettes are susceptible to humidity, only remove cuvettes for direct use and close canister immediately after!

ACCURATE & RELIABLE

  • Operating ambient temperature 15°C–40°C
  • Photometric azide methemoglobin method
  • Measuring range: 0–25.6 g/dL; 0–15.9 mmol/L
  • Imprecision: CV <2 % •    Linearity: 0–20 g/dL: ±0.3 g/dL; >20 g/dL: ±0.7 g/dL
  • Calibrated to NCCLS reference method
  • Control cuvette included
  • Stores 4,000 patient results
  • Connect to PC via USB connector cable