MultiCare Error Codes

Ethitech Multicare 1, Ethitech

Warning: Low Battery

At this time, battery is getting low but you can still perform about 10 tests.

Solution

Replace the battery soon.

Ethitech Multicare 2, Ethitech

Warning: Low Battery

Battery power is low.
If you press the Enter button after discharging the battery, the battery icon will flash and then after ten seconds the analyzer will turn off automatically.

Solution

Replace the battery immediately.

Error Messages

In certain circumstances error messages may appear on your display. Generally, you should first try the solutions suggested for the respective error. If the problem persists, please contact your local customer support and service center

Ethitech Multicare 3, Ethitech

Test Panel Error

Defective test panel or the test panel is damaged or inserted improperly.

Solution

Discard this test panel and test again using a new test panel.

Ethitech Multicare 4, Ethitech

Sample Error

An insufficient amount of sample volume was applied to the test panel.

Solution

Discard this test panel and test again using a new test panel. When testing again, apply the proper volume of sample mixture.

Ethitech Multicare 5, Ethitech

Expired Test Kit

The test kit date has expired.

Solution

Discard this test kit and test again using a new test kit that is not past the expiration date.

Ethitech Multicare 6, Ethitech

Temperature Error

If the environmental temperature is above or below the operating range, a thermometer icon will appear on the screen.

Solution

Move to an area with temperatures between the applicable temperature range, and perform a test. Do not artificially heat or cool the analyzer.

Ethitech Multicare 7, Ethitech

Communication Error

The communication between the analyzer and the printer failed.

Solution

Reconnect the analyzer and the mobile printer. If the error continues, turn the analyzer off. And then, turn on the analyzer again. If there is still an error message, please contact your local customer support and service center.

Ethitech Multicare 9, Ethitech

Extremely Total Hemoglobin

The applied sample is either under 7g/dL or above 23g/dL value of the total hemoglobin.

Solution

If this error occurs with a sample known to have total hemoglobin in the normal range, turn the analyzer off and back on and repeat test. If there is still an error message, please contact your local customer support and service center.

Ethitech Multicare 10, Ethitech

Communication Error

The code chip is broken. So, the analyzer could not communicate with the code chip.

Solution

Re-insert the code chip. If there is still an error message, please contact your local customer support and service center.

Ethitech Multicare 11, Ethitech

Internal Error Message for the Analyzer

It happens when an internal error occur.

Solution

Turn the analyzer off and then on again. If there is still this error message, please contact your local customer support and service center.