Afinion Error Codes

Information codes caused by test-specific limitations

hidden
103
Cause

Hemoglobin too low

Action to Take

Consult the Affinion HbA1c Package Insert

104
Cause

Hemoglobin too high

Action to Take

Consult the Affinion HbA1c Package Insert

105
Cause

HbA1c too low

Action to Take

Consult the Affinion HbA1c Package Insert

106
Cause

HbA1c too high

Action to Take

Consult the Affinion HbA1c Package Insert

107
Cause

Creatine too high

Action to Take

Consult the Affinion ACR Package Insert

108
Cause

Blood in urine

Action to Take

Consult the Affinion ACR Package Insert

Information codes caused by test-specific limitations

hidden
27
Cause

Start-up procedure failed

Action to Take

Contact your local supplier for assistance.

28
Cause

Start-up procedure failed

Action to Take

Contact your local supplier for assistance.

29
Cause

Start-up procedure failed

Action to Take

Contact your local supplier for assistance.

Self-test error. Analyzer in non-operative mode
Cause

Analyzer failure

Action to Take

Restart analyzer. If the problem persists, contact your local Afinion 2 supplier.

301
Cause

Self-test failed

Action to Take

Restart the analyzer.

302
Cause

Analyzer failure

Action to Take

Restart the analyzer and run controls. Repeat the test with a new sample and test cartridge.

303
Cause

Analyzer temperature is too high

Action to Take

Ensure that the operating temperature is within recommended range (15-32ºC, 59-89 °F). Wait until the analyzer has cooled down. Repeat the test with a new sample and test cartridge.

304
Cause

Analyzer temperature is too low

Action to Take

Ensure that the operating temperature is within recommended range for the Afinion Test in use (see package insert). The analyzer temperature is displayed in the Start-Up menu (see page 11). Repeat the test with a new sample and test cartridge.

305
Cause

Printer improperly connected Malfunction of the printer

Action to Take

Switch off the analyzer, reconnect the printer and restart the analyzer. If the message persists, see the printer user manual.

Touch screen error
Cause

Touch screen failure/ Touch screen buttons do not respond accurately

Action to Take

Restart analyzer and realign screen.

Information codes caused by sample or test cartridge

Hidden
201
Cause

Insufficient sample volume:

  • Empty capillary
  • Air bubble in capillary
  • Capillary incompletely filled
Action to Take

Repeat the test with a new sample and test cartridge.
Ensure that the capillary is completely filled with no air bubbles (see package insert for the Afinion Test in use).

202
Cause

Excess sample on the sampling device exterior

Action to Take

Repeat the test with a new sample and test cartridge. Ensure that only the tip of the capillary is in contact with the sample (see package insert for the Afinion Test in use).

203
Cause

Wrong sample material

Action to Take

Repeat the test with a new sample and test cartridge. Ensure that proper sample material is used (see package insert for the Afinion Test in use, section “Specimen collection and storage”).

204
Cause

Coagulated sample

Action to Take

Repeat the test with a new sample and test cartridge. The time from filling the capillary until analyzing the test cartridge should be as short as possible.

Cause

Hemolysed blood sample or poor sample quality

Action to Take

Consult the Afinion Package Insert. Repeat the test with a new sample and test cartridge.

Cause

Test cartridge or analyzer failure

Action to Take

Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.

205
Cause

Capillary cracked or damaged

Action to Take

Repeat the test with a new sample and test cartridge. Inspect the sampling device before use and handle with care.

206
Cause

Barcode label not readable (dirty or damaged)

Action to Take

Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.

207
Cause

No sampling device inserted
Sampling device belongs to another Afinion Test
Label on sampling device not readable (dirty or damaged)

Action to Take

Repeat the test with a new sample and test cartridge. Ensure that the correct sampling device is in place and that the sampling device label is clean

208
Cause

Test cartridge previously used

Action to Take

Repeat the test with a new sample and test cartridge.

209
Cause

Test cartridge has passed expiration date

Action to Take

Check expiry date on the foil pouch or kit container. Repeat the test using a new sample and a new test cartridge from another lot.

Cause

The date in the analyzer is incorrectly set

Action to Take

Check the date in the analyzer to make sure it is set correctly. Repeat the test with a new sample and test cartridge.

210
Cause

Test cartridge temperature too low

Action to Take

Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Afinion Test in use).

211
Cause

Test cartridge temperature too high

Action to Take

Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Afinion Test in use).

212
Cause

Software upgrade is required to run this test.

Action to Take

Contact your local supplier for assistance.

213
Cause

Test cartridge or analyzer failure

Action to Take

Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.

214
Cause

Test cartridge or analyzer failure

Action to Take

Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.

215
Cause

Test cartridge or analyzer failure

Action to Take

Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls

Cause

Hemolysed blood sample or poor sample quality (Afinion HbA1c)

Action to Take

Consult the Afinion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.

217
Cause

Hemolysed blood sample or poor sample quality (Afinion HbA1c)

Action to Take

Consult the Afinion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.

218
Cause

Condensation detected on cartridge

Action to Take

Run a new test cartridge. Ensure that the cartridge is equilibrated to room temperature before the foil pouch is opened.

Other Information Codes

hidden
401
Cause

No registered supervisors in operator list

Action to Take

At least one supervisor is required in the operator list when the analyzer is configured to operator ID verified, (see page 14).

402
Cause

Cannot delete last supervisor

Action to Take

At least one supervisor is required in the operator list when the analyzer is configured to operator ID verified, (see page 14).

403
Cause

This assay type is not accessible to the operator

Action to Take

The operator logged in does not have access to run this assay type. Please contact your supervisor.

404
Cause

Operator ID is not found in operator list

Action to Take

When Operator ID with verification is enabled, operator ID entered is required to be present in the operator list, (see page 14). Please contact your supervisor.

hidden
501
Cause

The control lot has passed expiration date

Action to Take

Check the expiration date on the control lot package insert or kit box. Repeat the test using a sample from a new control lot.

502
Cause

Afinion Control Data is not recognised and is not stored in control lot database

Action to Take

Reenter the Afinion Control Data, (see page 16).

503
Cause

Control verification aborted

Action to Take

The Afinion Control Data entered was not recognised. The control test was aborted by the operator. Test result was not stored. Run new control test to reset QC lockout interval.

504
Cause

Required control test interval has expired. Patient testing is disabled for this assay

Action to Take

A passed control run must be performed according to configuration to unlock this assay for patient testing

hidden
601
Cause

Operator list or control lot database is full

Action to Take

The operator list can store 1000 operators and the control lot database can store 100 control lots. Delete an operator or control lot to enter a new item.