Information codes caused by test-specific limitations
hidden
103
Cause
Hemoglobin too low
Action to Take
Consult the Affinion HbA1c Package Insert
104
Cause
Hemoglobin too high
Action to Take
Consult the Affinion HbA1c Package Insert
105
Cause
HbA1c too low
Action to Take
Consult the Affinion HbA1c Package Insert
106
Cause
HbA1c too high
Action to Take
Consult the Affinion HbA1c Package Insert
107
Cause
Creatine too high
Action to Take
Consult the Affinion ACR Package Insert
108
Cause
Blood in urine
Action to Take
Consult the Affinion ACR Package Insert
Information codes caused by test-specific limitations
hidden
27
Cause
Start-up procedure failed
Action to Take
Contact your local supplier for assistance.
28
Cause
Start-up procedure failed
Action to Take
Contact your local supplier for assistance.
29
Cause
Start-up procedure failed
Action to Take
Contact your local supplier for assistance.
Self-test error. Analyzer in non-operative mode
Cause
Analyzer failure
Action to Take
Restart analyzer. If the problem persists, contact your local Afinion 2 supplier.
301
Cause
Self-test failed
Action to Take
Restart the analyzer.
302
Cause
Analyzer failure
Action to Take
Restart the analyzer and run controls. Repeat the test with a new sample and test cartridge.
303
Cause
Analyzer temperature is too high
Action to Take
Ensure that the operating temperature is within recommended range (15-32ºC, 59-89 °F). Wait until the analyzer has cooled down. Repeat the test with a new sample and test cartridge.
304
Cause
Analyzer temperature is too low
Action to Take
Ensure that the operating temperature is within recommended range for the Afinion Test in use (see package insert). The analyzer temperature is displayed in the Start-Up menu (see page 11). Repeat the test with a new sample and test cartridge.
305
Cause
Printer improperly connected Malfunction of the printer
Action to Take
Switch off the analyzer, reconnect the printer and restart the analyzer. If the message persists, see the printer user manual.
Touch screen error
Cause
Touch screen failure/ Touch screen buttons do not respond accurately
Action to Take
Restart analyzer and realign screen.
Information codes caused by sample or test cartridge
Hidden
201
Cause
Insufficient sample volume:
- Empty capillary
- Air bubble in capillary
- Capillary incompletely filled
Action to Take
Repeat the test with a new sample and test cartridge.
Ensure that the capillary is completely filled with no air bubbles (see package insert for the Afinion Test in use).
202
Cause
Excess sample on the sampling device exterior
Action to Take
Repeat the test with a new sample and test cartridge. Ensure that only the tip of the capillary is in contact with the sample (see package insert for the Afinion Test in use).
203
Cause
Wrong sample material
Action to Take
Repeat the test with a new sample and test cartridge. Ensure that proper sample material is used (see package insert for the Afinion Test in use, section “Specimen collection and storage”).
204
Cause
Coagulated sample
Action to Take
Repeat the test with a new sample and test cartridge. The time from filling the capillary until analyzing the test cartridge should be as short as possible.
—
Cause
Hemolysed blood sample or poor sample quality
Action to Take
Consult the Afinion Package Insert. Repeat the test with a new sample and test cartridge.
—
Cause
Test cartridge or analyzer failure
Action to Take
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
205
Cause
Capillary cracked or damaged
Action to Take
Repeat the test with a new sample and test cartridge. Inspect the sampling device before use and handle with care.
206
Cause
Barcode label not readable (dirty or damaged)
Action to Take
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
207
Cause
No sampling device inserted
Sampling device belongs to another Afinion Test
Label on sampling device not readable (dirty or damaged)
Action to Take
Repeat the test with a new sample and test cartridge. Ensure that the correct sampling device is in place and that the sampling device label is clean
208
Cause
Test cartridge previously used
Action to Take
Repeat the test with a new sample and test cartridge.
209
Cause
Test cartridge has passed expiration date
Action to Take
Check expiry date on the foil pouch or kit container. Repeat the test using a new sample and a new test cartridge from another lot.
—
Cause
The date in the analyzer is incorrectly set
Action to Take
Check the date in the analyzer to make sure it is set correctly. Repeat the test with a new sample and test cartridge.
210
Cause
Test cartridge temperature too low
Action to Take
Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Afinion Test in use).
211
Cause
Test cartridge temperature too high
Action to Take
Repeat the test with a new sample and a new test cartridge. Ensure that the operating temperature is within acceptable range (see package insert for the Afinion Test in use).
212
Cause
Software upgrade is required to run this test.
Action to Take
Contact your local supplier for assistance.
213
Cause
Test cartridge or analyzer failure
Action to Take
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
214
Cause
Test cartridge or analyzer failure
Action to Take
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls.
215
Cause
Test cartridge or analyzer failure
Action to Take
Repeat the test with a new sample and test cartridge. If the problem persists, restart the analyzer and run controls
—
Cause
Hemolysed blood sample or poor sample quality (Afinion HbA1c)
Action to Take
Consult the Afinion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.
217
Cause
Hemolysed blood sample or poor sample quality (Afinion HbA1c)
Action to Take
Consult the Afinion HbA1c Package Insert. Repeat the test with a new sample and test cartridge.
218
Cause
Condensation detected on cartridge
Action to Take
Run a new test cartridge. Ensure that the cartridge is equilibrated to room temperature before the foil pouch is opened.
Other Information Codes
hidden
401
Cause
No registered supervisors in operator list
Action to Take
At least one supervisor is required in the operator list when the analyzer is configured to operator ID verified, (see page 14).
402
Cause
Cannot delete last supervisor
Action to Take
At least one supervisor is required in the operator list when the analyzer is configured to operator ID verified, (see page 14).
403
Cause
This assay type is not accessible to the operator
Action to Take
The operator logged in does not have access to run this assay type. Please contact your supervisor.
404
Cause
Operator ID is not found in operator list
Action to Take
When Operator ID with verification is enabled, operator ID entered is required to be present in the operator list, (see page 14). Please contact your supervisor.
hidden
501
Cause
The control lot has passed expiration date
Action to Take
Check the expiration date on the control lot package insert or kit box. Repeat the test using a sample from a new control lot.
502
Cause
Afinion Control Data is not recognised and is not stored in control lot database
Action to Take
Reenter the Afinion Control Data, (see page 16).
503
Cause
Control verification aborted
Action to Take
The Afinion Control Data entered was not recognised. The control test was aborted by the operator. Test result was not stored. Run new control test to reset QC lockout interval.
504
Cause
Required control test interval has expired. Patient testing is disabled for this assay
Action to Take
A passed control run must be performed according to configuration to unlock this assay for patient testing
hidden
601
Cause
Operator list or control lot database is full
Action to Take
The operator list can store 1000 operators and the control lot database can store 100 control lots. Delete an operator or control lot to enter a new item.